Case Report

Long-term outcomes of pediatric bilateral sequential auditory brainstem implant user: A case report

Volume 17 · Issue 3 Publish Date: July 15, 2021
Full Text PDF
DOI
Betül Çiçek Çınar
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Merve Özbal Batuk
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Mehmet Yaralı
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Gonca Sennaroğlu
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Esra Yücel
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Burçak Bilginer
Department of Neurosurgery, Hacettepe University, Medical Faculty, Ankara, Turkey
Levent Sennaroğlu
Department of Otolaryngology, Hacettepe University, Medical Faculty, Ankara, Turkey
Çınar, B. Çiçek, Özbal Batuk, M., Yaralı, M., Sennaroğlu, G., Yücel, E., Bilginer, B., & Sennaroğlu, L. (2021). Long-term outcomes of pediatric bilateral sequential auditory brainstem implant user: A case report. B-ENT, 17(3), 192–195. https://doi.org/10.5152/B-ENT.2021.20170
Full Text Full Text PDF

Abstract

Auditory brainstem implants (ABIs) are neuroprosthetic devices that stimulate the cochlear nuclei directly. Unilateral ABI applications are more common than bilateral. This paper reports our experience with a 7-year-old bilateral ABI user. Subjective and objective tests were applied, and bilateral profound sensorineural hearing loss was confirmed. The computerized tomography and magnetic resonance imaging of the temporal bone revealed bilateral cochlear aplasia with dilated vestibule. Intraoperative eABR was measured during both surgeries. Meaningful Auditory Integration Scale (MAIS), Categories of Auditory Performance (CAP) II, and Speech Intelligibility Rating (SIR) were used for evaluations of auditory perception. To evaluate auditory performance and hearing related quality of life, Functioning After Pediatric Cochlear Implantation instrument (FAPCI) and Pediatrics Quality of Life (PedsQL) scales were used. Free-field results improved rapidly following the second implant and soon reached 50–65 dB 0.5, 1, 2. and 4 kHz. With both implants, the CAP II and SIR scores were improved. ABI procedure is mostly performed unilaterally, but the outcomes tend to fall short of those attained by children with normal anatomy. Although our patient benefited significantly from her bilateral ABI, the surgery remains complex and is not routinely undertaken.

Similar Articles

Article Info
Published In
Journal B-ENT
Volume / Issue Volume 17 · Issue 3
Pages 192-195
History
Published Online July 15, 2021
Affiliations
Betül Çiçek Çınar
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Merve Özbal Batuk
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Mehmet Yaralı
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Gonca Sennaroğlu
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Esra Yücel
Department of Audiology, Hacettepe University, Health Science Faculty, Ankara, Turkey
Burçak Bilginer
Department of Neurosurgery, Hacettepe University, Medical Faculty, Ankara, Turkey
Levent Sennaroğlu
Department of Otolaryngology, Hacettepe University, Medical Faculty, Ankara, Turkey
Cite this Article
Çınar, B. Çiçek, Özbal Batuk, M., Yaralı, M., Sennaroğlu, G., Yücel, E., Bilginer, B., & Sennaroğlu, L. (2021). Long-term outcomes of pediatric bilateral sequential auditory brainstem implant user: A case report. B-ENT, 17(3), 192–195. https://doi.org/10.5152/B-ENT.2021.20170
Outlines